
This is a high concentration oxygen analyzer using limiting current method of Zirconia electrolyte for clean gas measurement.

| DEL | FG-IIUX-S | FG-IIU-100RS |
|---|---|---|
| Principle | Limiting current method of Zirconia electrolyte | |
| Installation | Indoor type、Plane・Wall type ※ Meter: Panel type | |
| Application | O2PSA、The oxygen enricher etc. | Synthetic air for the hospital etc. |
| Range | 0~95%O2 | 0~25%O2 |
| Display | 00.00~100.00%O2 | 00.00~25.00%O2 |
| Output | D.C.4-20mA to 0~100%O2 | D.C.4-20mA to 0~25%O2 |
| Alarm | Output alarm 2 points | |
| Linearity/Repeatability | Less than ±1%FS | |
| Response | 90% reading in 10 sec. (Swinging to a high density side) | |
| Drift | Less than ±2%/week | |
| Sampling | Introduce (positive pressure) | |
| Gas Pressure | 0.01~0.05MPa | |
| Gas temperature | 80℃max | |
| Piping | Inlet & Outlet PT1/4 | |
| Warm-up | About 20 minutes | |
| Power | A.C.100V ± 10V 50/60Hz 200VA max. | |
| Using Temp. | 0~50℃ | |
| Accessory | Connecting cord、Power cord 3m、Digital panel meter | |
| Option | Flow meter w | |
※About 2 years warranty
・It is limited our original instrument only.
・Consumption parts such as filter, O-ring are except
・In case of the sampling gas contains strong acidity, strong alkaline, or corrosiveness, it may be out of warranty.