TB seriesc

TB seriesBack to Product line page

TB sensor

*It is not possible to make a measurement only by using a sensor.Above control unit is necessary to measure oxygen.

TB-ⅡG

TB-ⅡG is an oxygen analyzer whose detective element is made of Zirconia electrolyte for measurement of exhaust gas.

Picture

Features
  • Several accessories with each object facility.
  • Built-in multi-filter protects detector from dirt and corrosion.
  • Simple structure, easy maintenance.
  • It is possible to measure CO2+H2 at the fuel surplus time.(display as "- O2%")
Applications
  • Boiler, Steel heating furnace, Glass fusion furnace, Ceramic sintering furnace, Waste heat plant,All burning exhaust gases.
Dimensions
Dimensions
The choice flow chart of the accessories
The choice flow chart of the accessories
Specifications
Measuring Range -20~25%O2
Linearity Less than ±1%FS(0-25%O2)
Repeatability Less than ±1%FS(0-25%O2)
Response 90% reading 10sec.(Swinging to a high density side)
Drift Less than±2%FS/week
Sample suction With Aspirator block, Guide probe
Sample pressure 0±5KPa(It depends on accessory)
Sample Temp. 300℃ MAX.(Sensor partonly)
0 ~ 700℃(In case of Guide Probe use)
0 ~1350℃(In case of Aspirator Block use)
Warm up time about 20 minutes
Option Rain protector

Accessories for TB-ⅡG

Aspirator block

The aspirator block is a suction equipment of the sample gas.
It combines with the sampling probe as the sample introduction pipe.
A direct measurement is possible under the high temperature atmosphere because it works by air introduction.


Picture

Dimensions
Dimensions
Specifications
Material Return type:SUS316L / Release to atmosphere type:SUS316
Installation Return type:JIS5K65A~ / Release to atmosphere type:JIS5K40A~
Sample gas temp. It follows to the specification of probe
Sample gas pressure Return type:±5kPa / Release to atmosphere type:-0.1kPa~1kPa
Tubing Rc 1/4
Guide probe

The guide probe is a sampling equipment which uses a flow of sampling gas.
It is installed at the place which the sample gas flow exists such a as gas duct.
The utility of instrumenting air and the power and so on is unnecessary because of sampling which used a sample gas stream.

Picture

Dimensions
Dimensions
Specifications
Material Low dust type:SUS316 / High dust type:SUS316L
Installation Low dust type:JIS5K50A~ / High dust type:JIS5K80A~
Length of insetion As your specification (MAX.1500㎜)
Sample gas temp. 700℃MAX
Sample gas pressure ±5kPa
Tubing Rc 1/4
TOP
TB-ⅡF

TB-ⅡF series is an oxygen analyzer whose detective element is made of Zirconia electrolyte for measurement of low concentration of oxygen.

Picture

Features
  • Wide range application, from 1ppm to 100%O2.
  • Built-in multi-filter protects detector from dirt and corrosion.
  • Accepts 0.2 to 5 L/min flow rate.
  • Simple structure, easy maintenance.
Dimensions
Dimensions
フローシート
フローシート
Specifications
MODEL TB-IIF-P Rc 1/8 female connector
TB-IIF-R Rc 1/4 female connector
TB-IIF-S O.D. 3/8" Swagelok
TB-IIF-V 4VCR fitting
Measuring Range 1ppm ~ 100%O2 / 1 ~ 10^-30 atmO2
Linearity The Larger either of less than ±1%FS or±1ppm
Repeatability The Larger either of less than ±1%FS or±1ppm
Response 90% reading 10sec.(Swinging to a high density side)
Drift Less than ±2%FS/week
Sample Temp. 300℃MAX
Sample Flow Rate 0.2~2L/min(TB-IIF-P)、others 0.5~5L/min
Sample pressure TB-IIF-V :203kPa~1.33×10-3Pa(2atm~10-5Torr)
Others:203kPa~1.33×10-1Pa(2atm~10-3Torr)
Warm up time About 20 minutes
TB-ⅡV TB-ⅡVN

TB-ⅡV series is an oxygen analyzer whose detective element is made of Zirconia electrolyte for measurement of small amount of oxygen in vacuum furnace.

Picture

Features
  • Direct installation for sealed vessel or vacuum furnace.
  • Measuring vacuum atmosphere under the pressure of 1×10-3Pa(1×10-5Torr).
  • Optional mounting flange is available.
TB-ⅡV Dimensions
TB-ⅡV Dimensions
TB-ⅡVN Dimensions
TB-ⅡVN Dimensions
TB-ⅡV Flow diagram
TB-ⅡV Flow diagram
TB-ⅡVN Flow diagram
TB-ⅡVN Flow diagram
Specifications
Flange size VF25, NW40, NW25 etc. (Please feel free to contact for other sizes)
Installation Vertical ~ Horizontal, above vacuum flange(Customer preparation)
Calibration gas inlet SUS TP O.D.1/4(O-ring fitting)
Measuring range 1ppm~100%O2 1~1×10-30atm O2
Linearity Large one either of less than ±1%FS or±1ppm
Repeatability Large one either of less than ±1%FS or±1ppm
Response 90% reading 10sec.(Swinging to a high density side)
Drift Less than ±2%FS/wk
Sampling Diffusion
Sample Temp. 300℃MAX
Sample pressure 2atm~1×10-3Pa(1×10-5Torr)
Warm up time About 20 minutes
Option The port connection for calibration
TOP
Click here to "Serch applications"

Control unit

Control unit C-28C C-48

These instruments should be combined with O2 sensor then control cell temperature and convert cell EMF to O2 value for display and signal output.

Picture

Features
  • Compact DIN size. (96mm×96mm)
  • Readable large display
  • Free-range
  • Serial communication function (RS-232C) is normally equipped.
  • Alarm contact A output(H, H/H, L, L/L) possible
  • Failure alarm (Sensor abnormal temperature failure)
Dimensions
Dimensions
Specifications
C-28C C-48
Dimensions 96W×96H×132D
Weight about 0.5kg
Power A.C.90 ~130V
Display 4 figures LED 5 figures LED
Disply Ranges R1:% 0~99.99、0~100.0
-19.99~99.99
R1 %:0.00~100.00
ppm:0.0~9999.9
atm:-35.00~0.00
mV:-40.0~1500.0
R2:ppm 0~9999、0~999.9 R2
R3:mV -525~1575mV R3
Output
※1
D.C.4~20mA
D.C.0 ~ 1V or 5V or 10V
Communication RS232C
Range Select Manual / Auto Manual / Auto / Remote
Alarm Output:Hi/Lo、Hi/HHi、Lo/LLo
Failure:Sensor temp. abnormal
Other Functions Cell EMF and/or Cell temp. are shown on lower display.
Range answer-back, hold, one-touch calibration equipped.
Control unit C-201W

These instruments should be combined with O2 sensor then control cell temperature and convert cell EMF to O2 value for display and signal output. This is the special model for foreign country that simplified the specification and made reasonable price.

Picture

Features
  • Reasonable price.
  • Easy to handle.
  • Readable large display.
  • Serial communication function (RS232C) is normally equipped.
Dimensions
Dimensions
Specifications
  MODEL:C-201W
Dimensions 215W×292H×95D
Weight Approx. 2500g
Power 85~264VAC
Display Upper 4 figures LED
Lower LCD
Display Ranges R1:% 0~99.99%
R2:ppm 0~9999ppm
Output
※1
DCA D.C.4~20mADC/0-1VDC/0-5VDC/0-10VDC (select one of them)
Communication RS232C
Range Select Manual/Auto/Remote
Alarm Output:Hi/Lo、Hi/H・Hi、Lo/L・Lo
Failure:Sensor temp. abnormal
(Either of open collectors)
Other functions Cell EMF and/or Cell temp. are shown on lower display.
Range answer-back, hold, one-touch calibration equipped.
TOP
Click here to "Serch applications"

TB-FI series

Portable oxygen analyzer TB-FI series

TB-FI series is an oxygen analyzer whose detective element is made of Zirconia electrolyte. TB-IIF-P sensor and C-101 control unit are installed altogether in a box for portable uses.

Picture

特徴
  • Wide range application, from 1ppm to 100%O2.
  • Built-in multi-filter protects detector from dirt and corrosion.
  • Output range can select freely.
  • Serial communication function is equipped normally.
  • AC85V~260V power supply can be applied.
Dimensions TB-FI
Dimensions TB-FI
Dimensions TB-FI
Dimensions TB-FI
Dimensions TB-FI-PC
Dimensions TB-FI-PC
Dimensions TB-FI-PR
Dimensions TB-FI-PR
Specifications
Principle Zirconia electro-chemical cell
Structure SS box, TB-ⅡP sensor, C-101 control unit, needle valve, flow meter etc.
Model TB-FI Pressurized sample
TB-FI-P Sampling pump installed
TB-FI-PC Sampling pump and charcoal filter installed
TB-FI-PR Sampling pump and charcoal filter and mist-separator installed
Measuring range 1ppm~100%O2
Display %:0~99.99%O2
ppm:0~9999ppmO2
Output D.C.4~20mA、0~1V or 0~10V(isolated)
F.S. can be set within the range of the above-mentioned display freely.
RS232C
Range change Auto/Manual
Initial setting 0~25%O2/0~1000ppmO2
Alarm Output :Hi/Lo、HHi/Hi or Lo/LLo(each A Point of contact・LCD Display)
Failure:Disconnection of Heater, RTD(LCD Display)
Linearity Large one either of less than ±1%FS or±1ppm
Repeatability Large one either of less than ±1%FS or±1ppm
Response 90% reading 10sec.(Swinging to a high density side)
Drift Less than ±2%FS/wk
Piping Rc1/4
Sample Flow Rate 0.2~2L/min
Sample Temp. 80℃ MAX.
Warm up time About 20 minutes
AC Power AC85V~AC260V
Accessory Power supply code
TOP
Click here to "Serch applications"

TB-WI

TB-WI

TB-WI series is an oxygen analyzer whose detective element is made of Zirconia electrolyte. TB-IIF-P sensor and C-201W control unit are installed altogether in a box for portable uses. This model is carried out reduction in cost and contributes as a built-in oxygen meter for foreign countries.

Picture

特徴
  • Reasonable price
  • Wide range application, from 1ppm to 100%O2.
  • Output range can select freely.
  • Serial communication function is equipped normally.
  • 85V~264VAC power supply can be applied.
Dimensions
Dimensions
Specifications
Principle Zirconia electro-chemical cell
Structure SS box, TB-ⅡP sensor, C-201W control unit, pump, needle valve, flow meter etc.
Measuring range 1ppm~100%O2
Display %:0~99.99%O2
ppm:0~9999ppmO2
Output 4~20mADC/0-1VDC/0-5VDC/0-10VDC (Select the one of them)
RS-232C
Range change Auto/Manual/Remote
Initial setting 0~25%O2/0~1000ppmO2
Alarm Hi/Lo、Hi/H・Hi、Lo/L・Lo
Failure:Disconnection of Heater, RTD
(All of them are open collector output)
Linearity The Larger either of less than ±1%FS or±1ppm
Repeatability The Larger either of less than ±1%FS or±1ppm
Response 90% reading 5sec.(Swinging to a high density side)
Drift Less than ±2%FS/wk
Piping Rc1/4
Sampling Supplying sample gas or vacuuming by inside pump
Sample Flow Rate 0.2~2L/min
Sample Temp. 100℃
Warm up time About 10min.
AC Power 85~264VAC
Accessory Power supply code
TOP
Click here to "Serch applications"
TOPPAGE